The Food and Drug Administration on Wednesday issued an early alert to notify the public that ICU Medical has recommended the removal of certain Portex Endotracheal Tubes. The agency described the situation as a potentially high-risk device issue.

ICU Medical identified a manufacturing issue affecting certain lots of the tubes, according to the FDA. The affected tubes were made using a resin with variability in the percentage of plasticizer, which may cause the tube material to be softer than expected.

Affected Portex Endotracheal Tubes
Item NumberDescriptionUnique Device Identifier
100/127/025TRACHEAL TUBE,UNCUFFED 2.5MM MURPHY EYE ORAL/NASAL 10/CA15019315019807
100/127/030TRACHEAL TUBE,UNCUFFED 3.0MM MURPHY EYE ORAL/NASAL 10/CA15019315019814
100/127/035TRACHEAL TUBE,UNCUFFED 3.5MM MURPHY EYE ORAL/NASAL 10/CA15019315019821
100/134/030TRACHEAL TUBE POLAR CLEAR SOUTH ORAL UNCUFFED 3.0MM 10/1501931502055
100/134/035TRACHEAL TUBE POLAR CLEAR SOUTH ORAL UNCUFFED 3.5MM 10/B15019315020568
Source: U.S. Food and Drug Administration.

The tubes are intended for oral and/or nasal intubation for airway management. The FDA said that using the affected tubes intended for the neonatal and early infant population may lead to a delay in intubation because the softness can increase resistance to ventilation, cause tube kinking, and make intubation difficult.

If medical staff do not notice the problem during intubation, the issue may result in inadequate ventilation and unscheduled tube replacement procedures, the agency said. These complications may lead to or contribute to serious adverse events including death. ICU Medical has reported one death associated with the issue as of Nov. 16.

ICU Medical sent a letter to affected customers on Nov. 20 recommending they check all inventory and immediately discontinue use of the products. The company recommended that facilities destroy the affected products or segregate and quarantine them if immediate destruction is not possible.

  1. Early Alert: ICU Medical Removes Portex Endotracheal Tubes — U.S. Food and Drug Administration
How this article was made
Written with
Gemini 3.1 Pro (Google)
Fact-checked by
Claude Sonnet 5.5 (Anthropic) — passed after revision
Generated
Nov. 26, 2026, 9:39 a.m. ET
Sources used
1 official source, listed above
Article type
Standard

The AI wrote only from the sources listed. No human reporter was involved. Read our standards or compare the models in the AI Report Card.